DIY-Thermocam absolute temperature accuracy

I'm using using the thermocam for thermal analysis of residential buildings. My application requires an overall accuracy of absolute temperature of about 1K. I found it difficult to achieve this with the builtin automatic calibration procedure. This article documents my experiments and the workflow I came up with.

The basic setup I'm working with is indoor measurement with wet ice as reference body. IE most objects are in the range 0°C - 25°C and the ambient temperature (both air and radiation) is about 20°C. At the start, end and several times in between I take a thermal image of the reference body. Such a campaign typically take about 20-30 minutes.

Automatic calibration performance

While working with the built-in calibration procedure, I noticed several issues:

  • The resulting calibration data is not very reproducible.
  • The slope of the calibration curve generally seems too low: Most of the time I get values around 30mK while the data sheet specifies <50mK. (Of course 30mK < 50mK actually is true, but it seems a bit off to me.)
  • The spot sensor, supposed to help calibration, drifts a lot over half an hour. Probably due to warming up.

The following table shows repeated measurement results for a bowl of wet snow:

Time	  mean raw value	temperature [°C]	spot sensor [°C]
------------------------------------------------------------------------
5:57:41	   7556,28		 4,03			 1,16
6:01:22	   7279,84		-3,95			-3,24
6:02:53	   7469,17		 0,97			-4,14
6:04:10	   7441,50		 4,36			-4,30
6:05:19	   7421,88		 2,86			-4,86
6:08:15	   7409,88		-1,21			-5,28
6:11:54	   7397,90		 5,38			-5,62
6:15:53	   7371,48		-6,23			-8,60
6:20:37	   7349,74		 0,81			-9,50
				===============
			average: 0,78 °C

While the average value of temperatures is quite close to the real value of 0°C, the individual temperatures derived from the thermal images are quite a bit off. The spot sensor gives a fairly decent temperature at the beginning, after switching the camera on. But then the values start running away.

Manual calibration

The most simple conversation formula from raw values to temperatures is

temperature = reference_temperature + (value - offset) * slope

To calculate slope two reference bodies are used: The bowl of wet ice (reference_temperature = 0°C) and an object in thermal equilibrium with the air in the room. A piece of dry cloth hanging on a hook from the ceiling is a good candidate.

The following image shows the average raw values of two reference bodies in such an experiment. The colder object is wet ice and then other is at 21.6°C. As you can see, the offset has quite some drift, while the difference between reference bodies (related to slope) is quite constant:

The data was extracted using the following gmic command line in a semi-manual way. For each image on the command line, the user is prompted to draw a rectangle of which gmic prints the average value to stdout:

gmic *_object_raw.pgm -crop -repeat '$!' '-l[$>]' -echo_stdout '{0,b}':'{is#0/wh#0}' -endl -done

Actually using linear interpolation the slope can be calculated at each measurement of either reference body. The next plot shows these values for slope together with a linear regression fit. As you can see, there is some trend for slope to become higher over time, but it is mostly within noise level.

Most slope values are within 5% of the mean. Thus when the reference temperature is 0°C, the error at 20°C is about 1K. It is possible to meet the requirements for my application without accounting for the drift in slope. I still take measurements of warm reference bodies at regularly, but only for monitoring measurement performance, not for calibration.

Also slope values around 47mK are very compatible with the data sheet specification of <50mK - so this method of calibration seems quite reasonable.