Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

gdx.start() with different period #49

Open
mctosima opened this issue Nov 25, 2022 · 3 comments
Open

gdx.start() with different period #49

mctosima opened this issue Nov 25, 2022 · 3 comments

Comments

@mctosima
Copy link

mctosima commented Nov 25, 2022

Hi, I'm using the GoDirect Respiratory Belt and EKG. I want to use both in different periods (sampling rate). I want to use the ECG at 40hz while it cant because the Respiratory Belt only supports 20hz. Is there any way to do this?

My capture code

@sswartley
Copy link
Contributor

Hello,

This does work using our Graphical Analysis app, so that is an option. That would be a way to get the data. But you are correct that our Python code is not able to handle this. I'm really not sure if you could dig into the Python code to get this to work. Sorry that I don't have any more insight into how this might work using our Python library.

All the best,

Sam

@mctosima
Copy link
Author

Hi @sswartley, thanks for replying. I agree that we can do that using the graphical analysis app. Unfortunately, the main feature that I need is missing on the Graphical Analysis app, which is set the scheduler to start at a certain time. That's why I did this on Python.

As you clarify that this individual sampling rate isn't available on the python packages, for now, I'm using the ECG with a USB while the RB with a BLE, and triggers 2 different capture codes from the terminal. I think this is the only solution for now. Thank you.

You may close this issue or leave it opened.

@sswartley
Copy link
Contributor

One thing that you might pursue is working with the godirect library directly, and not using the 'gdx' module. If you take a look at the examples you will find one that does not use the 'gdx' module. This example sends back the sensor values as a list. I do not know for sure, but maybe this example (modified to work with 2 Go Direct devices) handles one sensor sending back a list that has more values than the other sensor's list.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment