SAR calculation depends on the duration of the event following the RF event #223

Open
h3lg3 opened this issue Dec 18, 2024 · 0 comments · May be fixed by #249
Open

SAR calculation depends on the duration of the event following the RF event #223

h3lg3 opened this issue Dec 18, 2024 · 0 comments · May be fixed by #249
Assignees

Comments

@h3lg3
Copy link
Contributor

h3lg3 commented Dec 18, 2024

Describe the bug
Changing the sampling duration in the example sequence write_epi.py affects the SAR calculation. Since TE and TR are fixed for the sequence, changing the sampling duration does not affect the sequence duration and hence should not affect SAR calculation.

To Reproduce
Run the write_epi.py script and append pp.SAR.SAR_calc.calc_SAR(seq) at the end. Now change the sampling duration and re-run the script.

sampling_time = 6.4e-3/2

https://github.com/imr-framework/pypulseq/blob/dev/examples/scripts/write_epi.py

Expected behavior
SAR calculation should be independent on other sequence events that are not RF related and that do not change the timing of the sequence.

Additional context
What happens in the write_epi.py is that changing the sampling duration affects the spoiler duration t_sp = 0.5 * (TE - readout_time - t_refwd). The SAR calculation is affected by the duration of the event following the RF event. In this case that is the duration of the spoiler. I think there is a bug in the SAR calculation. In Line 101 and further in SAR_calc.py it reads:

  for block_counter in block_events:
      block = seq.get_block(block_counter)
      block_dur = calc_duration(block)
      t[block_counter - 1] = t_prev + block_dur
      t_prev = t[block_counter - 1]
      if hasattr(block, "rf"):  # has rf
          rf = block.rf
          signal = rf.signal
          # This rf could be parallel transmit as well
          SAR_wbg[block_counter] = _calc_SAR(Qtmf, signal)
          SAR_hg[block_counter] = _calc_SAR(Qhmf, signal)

The index of the time vector t is decremented by 1, however the decrement is missing for the SAR_wbg vector, which leads to a misalignment of time and SAR.

Here you can see the sequence diagram. Please be aware that I deactivated the dummy scans in write_epi.py for sake of clarity.
image

And here is the corresponding time and SAR vector. The SAR for the 180 refocussing pulse is associated with a time stamp of 11.99 ms which is the end of the spoiler and not the end of the RF pulse.

IDX t[ms] RF
1: 0.25 0.0
2: 2.95 0.0
3: 6.5 16.5908
4: 8.7 0.0
5: 11.99 103.6925
6: 15.21 0.0
7: 18.5 0.0
8: 20.7 0.0

In my opinion the SAR vector should be decremented according to the time vector.

  for block_counter in block_events:
      block = seq.get_block(block_counter)
      block_dur = calc_duration(block)
      t[block_counter - 1] = t_prev + block_dur
      t_prev = t[block_counter - 1]
      if hasattr(block, "rf"):  # has rf
          rf = block.rf
          signal = rf.signal
          # This rf could be parallel transmit as well
          SAR_wbg[block_counter - 1] = _calc_SAR(Qtmf, signal)
          SAR_hg[block_counter - 1] = _calc_SAR(Qhmf, signal)

Thanks a lot for your support! I hope I understood the issue right.
Kind regards
Helge

h3lg3 added a commit to h3lg3/pypulseq that referenced this issue Feb 12, 2025
@schuenke schuenke linked a pull request Feb 12, 2025 that will close this issue